Modules / Nonconformity & CAPA / CAPA-1042 Search the system… Ctrl K
Viewing as

CAPA-1042 — Sterilization load release

عدم المطابقة والإجراءات
Investigation, root cause, actions and verification for this case.
CAPA-1042CriticalVerification SLA 17d
DetailsRoot CauseAction PlanEffectiveness VerificationLinksAudit Trail

Record

FieldValue
SourceAudit · AUD-2026-004 / F-03
SeverityCritical
EntityBlood Bank Centre
Opened2026-11-12 · by Audit Director
SLA due2026-11-30 · 17 days
StatusVerification

SLA & escalation

17 days to SLA

5 Whys chain builder

Why 1 — load released without Bowie-Dick record?
Record was on paper log, machine offline
Why 2 — machine offline?
Data module not connected to hospital network
Why 3 — not connected?
IT request pending since Sep (ticket 8841)
+ Add why…
chain continues until systemic cause

Fishbone categories

Equipment ● selectedMethodPeopleEnvironment
Primary category: Equipment · contributing: Method (paper log allowed)

Action plan

TypeActionOwnerDueEvidenceStatus
ImmediateQuarantine loads released 20–22 OctDr. N. Ali2026-11-142 filesDone
CorrectiveConnect sterilizer data module to network (IT-8841)IT + BB Dept.2026-11-201 fileIn progress
CorrectiveRetrain staff on load release checklistBB Supervisor2026-11-22—Planned
PreventiveElectronic interlock: no release without recordBiomed Eng.2026-12-15—Planned

Verification checklist

CheckResultComment
No repeat findings in 30-day windowPass0 repeats
Electronic interlock livePendingdue 15 Dec
Staff re-training completedPendingscheduled 22 Nov

Graph view — linked objects

MOH-SOP-0154 policyCBAHI MMS.02 standard AUD-2026-004 auditRS-018 risk CRS-77 training

Immutable audit trail

TimestampActorActionDetail
2026-11-12 14:02Audit DirectorcreateNC raised from F-03 (critical)
2026-11-12 14:20CAPA Quality Manageraccepttriage accepted · SLA set 2026-11-30
2026-11-13 08:05System (SoD)sod_blockaction owner cannot verify effectiveness